XT-100
Our most versatile product with a single input channel and a customizable delay range
Recommended if you need a finer step resolution and broad selection of delay ranges
XT-100 Overview
The latest generation of the product that started it all. The XT-100 features Colby's patented trombone technology for unmatched precision and ease of use. Using delay extensions, the XT-100 also comes in a number of configurations that give you the range you need, up to 100 ns.
Single channel offers 0-625 ps of total delay
Additional configurations with delay extensions offer total delay up to 2.5 ns, 5.0 ns, 10 ns, 20 ns, 50 ns, 80 ns, or 100 ns
Full wideband passive delay from DC to 18 GHz
Resolution step size as small as 0.25 ps
Trombone mean step accuracy better than 0.05% σ = 0.05 ps
500,000 operations before recommended maintenance service
Now with web browser interface for remote instrument control
XT-100 Technical Specs
XT-100 | |
---|---|
Technology Type | Trombone, Trombone + Relays |
# of Channels | 1 |
Signal Input Range | 0 - 18.0 GHz |
Min. Step Resolution | 0.25 ps |
Max Delay Range | 625 ps, 2.50 ns, 5.00 ns, 10.0 ns, 20.0 ns, 50.0 ns, 80.0 ns, 100.0 ns |
Phase Shift Step Resolution | 0.18 ° per 1 GHz |
Total Phase Shift at 1 GHz | 225°, 900°, 1800°, 3600°, 7200°, 18000°, 28800°, 36000° |
Total Phase Shift at 5 GHz | 1125°,4500°, 9000°, 18000°, 36000°, 90000°, 144000°, 180000° |
Switching Speed* | 250 ms - 6500 ms |
Ext. Trigger | no |
Ethernet TCP/IP | yes |
Serial RS-232 | yes |
MT-100A | optional |
Web Browser UI | yes |
Microwave Relays Rated | 5m MTBF |
Recommended Service Interval | 500,000 operations or 1 year |
Min. frequency for 360° phase shift coverage | 1.6 GHz, 400 MHz, 200 MHz, 100 MHz, 50 MHz, 20 MHz, 12.5 MHz, 10 MHZ |
Dimensions | 12" L x 16 3/4" W x 3 1/2" H (2U) |
Weight | 4.1 kg (9.0 lbs.) to 5.0 kg (11 lbs.) |
*depending on network latency
Optional Add On
MT-100A MICROTERMINAL
MT-100A provides convenient LCD display and numeric keypad entry for manual control of desired delay.
-
XT-100 Data Sheet
-
XT-100 Insertion/Return Loss Report
-
XT-100 User Manual